(1)    Exam Light Override

Allows desk controls to work properly when an indirect
ophthalmoscope is not installed.

(2)    Ophthalmoscope
        Override

Overrides the rheostat to provide full power illumination to Ophthalmoscope.

(3)    Ophthalmoscope
          Rheostat

Provides adjustable illumination power to the ophthalmoscope.

(4)    Ophthalmoscope
        Hanger

Provides storage for indirect ophthalmoscope and switches power on/off.  Lifting the ophthalmoscope dims desk and room lights to levels pre-set at dimmer (2) on instrument well panel.

(5)    Ophthalmoscope
          Power Receptacle

Twist lock receptacle compatible with most indirect ophthalmoscopes.

(6)    Fuse

Replaceable fuse.  6, 7.5V use 3 amp fuse - 12V use 1 amp fuse.

Voltage Selection

Output voltage of the Indirect Control Center is factory set.  To change the voltage to be compatible with an indirect ophthalmoscope using another voltage, follow the instructions below.

Remove trial lens drawer by pressing up on levers and lifting the drawer off the drawer guides

Locate terminal block below indirect control center and remove black wire from terminal block.

Locate desired voltage connection using the chart below.  Connect the black wire to the connector that corresponds to the desired voltage.

Replace trial lens drawer.

Color Code For Wires To Connectors
Violet = 6.5 VAC
Yellow = 7.5 VAC
Blue = 12 VAC
Orange = control wire
White = 0 VAC (common)

Return To Home Page.
Take Me Home!!

NOTE:  Changing to or from 12 volts requires a different rheostat.
              Check with factory for details.